Page 32 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01
mmmmm
Operation
Example
P054 provides up to 32 breakpoint levels used with P055 (Volume Breakpoint) for universal
volume calculation. The illustration indicates how you can set secondary indexes to
specific functions.
Starting Measurement
Startup
The HydroRanger 200 starts in DISTANCE mode with the transducer preset for the XPS-10
and an empty distance of 5 meters. Change the following parameters to reflect your
application parameters.
Parameter Index Value Description
P001 G 1 Operation = level
P002 G 1 Material = liquid
P003 G 2 Maximum Process Speed = medium
P004 G 104 Transducer = XPS-15
P005 G 1 Units = meters
P006 G 12 Empty = 12m
P007 G 10 Span = 10m
1
P054
[6 = 2m]
[3 = 1.5m]
[1 = 0.75m]
A
B
C
A
P054 [1.6] = 2m
B
P054 [1,3] = 1.5m
C
P054 [1,1] = .75m