viiimmmmmTable of ContentsP622 Inflow / Discharge Adjust...
Page 86 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FCO1mmmmmCommunications InstallationCommunications InstallationWiring Guidelines• the RJ-11 cable
7ML19981FCO1 HydroRanger 200 – INSTRUCTION MANUAL Page 87mmmmmCommunications InstallationPort 1: RS-232 RJ-11 Jack To connect the unit to a PC using a
Page 88 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FCO1mmmmmCommunications InstallationTo connect the unit to a modem using an RS-485 port:Configuri
7ML19981FCO1 HydroRanger 200 – INSTRUCTION MANUAL Page 89mmmmmCommunications InstallationP772 Baud RateThe communication rate with the master device.T
Page 90 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FCO1mmmmmCommunications InstallationP778 Modem AvailableSets the HydroRanger 200 to use an extern
7ML19981FCO1 HydroRanger 200 – INSTRUCTION MANUAL Page 91mmmmmModbus Register TypesModbus Register MapFeatures affecting volume, mA input, and Average
Page 92 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FCO1mmmmmModbus Register TypesThe HydroRanger 200 was designed to make it easy for master devices
7ML19981FCO1 HydroRanger 200 – INSTRUCTION MANUAL Page 93mmmmmModbus Register TypesProduct ID (R40,064)Identifying Milltronics device type:Point Data
Page 94 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FCO1mmmmmModbus Register TypesDiscrete Inputs (R41,070) This table shows the current status of th
7ML19981FCO1 HydroRanger 200 – INSTRUCTION MANUAL Page 95mmmmmModbus Register TypesPump OFF Setpoint (R41,430 – R41,435) The OFF setpoint level (P113)
ixmmmmmTable of ContentsP773 Parity ...
Page 96 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FCO1mmmmmModbus Register TypesUsually, the parameters are all read / write.Each parameter registe
7ML19981FCO1 HydroRanger 200 – INSTRUCTION MANUAL Page 97mmmmmModbus Register TypesGlobal Index Method (P782 = 0)Global format method sets index value
Page 98 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FCO1mmmmmModbus Register Types2. Read the value from the appropriate parameter register (the exam
7ML19981FCO1 HydroRanger 200 – INSTRUCTION MANUAL Page 99mmmmmModbus Register TypesGlobal Index Method (P782 = 0)Only the final digit of the format wo
Page 100 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FCO1mmmmmModbus Register TypesExamples of using the format word for both the index values and th
7ML19981FCO1 HydroRanger 200 – INSTRUCTION MANUAL Page 101mmmmmData TypesData TypesThe HydroRanger 200 parameters do not always use integers to hold v
Page 102 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FCO1mmmmmData TypesFor example, if R41,442 is read as a UINT32, the 32 bits would look like this
7ML19981FCO1 HydroRanger 200 – INSTRUCTION MANUAL Page 103mmmmmData TypesText MessagesIf a Milltronics device parameter returns a text message, that m
Page 104 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FCO1mmmmmData TypesRelay Function Codes (P111 Only) If a Milltronics device parameter returns a
7ML19981FCO1 HydroRanger 200 – INSTRUCTION MANUAL Page 105mmmmmError HandlingError HandlingModbus ResponsesWhen polled by a Modbus Master, a slave dev
xmmmmmTable of ContentsP901 Memory ...
Page 106 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FCO1mmmmmError Handling• If P000 is activated, then the value will be ignored and no error respo
7ML19981FCO1 HydroRanger 200 – INSTRUCTION MANUAL Page 107mmmmmError HandlingCommunication TroubleshootingGenerally1. Check the following:• There is p
Page 108 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FCO1mmmmmCommunication Appendix ACommunication Appendix A: Single Parameter Access (SPA)This App
7ML19981FCO1 HydroRanger 200 – INSTRUCTION MANUAL Page 109mmmmmCommunication Appendix AWriting ParametersTo set parameters through Modbus do the follo
Page 110 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FCO1mmmmmCommunication Appendix AError CodesThe error codes returned in the format area are 8-bi
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 111mmmmmParametersParameter ReferenceThe HydroRanger 200 is configured through its parameters,
Page 112 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersNotes
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 113mmmmmParametersP000 LockSecures the HydroRanger 200 from changes. Access this parameter dire
Page 114 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersFor these operations any of three level points (transducer 1, transducer 2, o
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 115mmmmmParametersP004 TransducerSpecifies the Siemens Milltronics transducer connected to the
ximmmmmTable of ContentsObstructions in the Sound Beam ... 223
Page 116 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP006 EmptyEnter distance in units (P005) from the face of the transducer to t
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 117mmmmmParametersAlways prevent the monitored surface from approaching within 0.33m (1 ft) of
Page 118 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersValues4 Half Sphere Bottom5 Flat Sloped Bottom6Flat Ends7 Parabola Ends8 Sphe
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 119mmmmmParametersP051 Maximum Volume For Readings in volumetric units (rather than percent), e
Page 120 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP054 Level Breakpoints (Universal Volume Calculation)When the tank shape is t
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 121mmmmmParametersEntering a Volume Breakpoint1. Go to Parameter P055.2. For each index enter a
Page 122 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersExamples: • If the measured value is in feet, enter 0.3333 to display the num
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 123mmmmmParametersP065 Reading Override ValueThis value is substituted for the current reading
Page 124 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersFailsafe (P070 to P072) P070 Failsafe TimerThe time for invalid measurements
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 125mmmmmParametersSelecting HI, LO, or HOLd1. Press FUNCTION to display the Auxiliary Functio
Page 126 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersRelays (P100 to P119)The HydroRanger 200 has six relays (or digital outputs)
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 127mmmmmParametersP110 Level SourceSets the level source on which the indexed relay is assigned
Page 128 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP112 Relay ON SetpointSets the process point at which the relay changes from
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 129mmmmmParametersP113 Relay OFF SetpointSets the process point at which the relay returns to i
Page 130 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP118 Relay Output LogicThe logic applied to relays to determine the contact o
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 131mmmmmParametersP119 Relay Logic TestForces the relay control logic into an ACTIVATED or DE-A
Page 132 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP122 Pump Service Ratio Selects pump usage based on the RUN time ratio rather
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 133mmmmmParametersTo select an independent Relay Failsafe value:1. Press FUNCTION to display
Page 134 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP131 Pump Run-On Duration Sets the number of seconds that the pump runs on.Yo
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 135mmmmmParametersP133 Pump Power Resumption Delay Sets the minimum delay before the first pump
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 1mmmmmHydroRanger 200HydroRanger 200The HydroRanger 200 is a single-point, six relay device tha
Page 136 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP137 Pump Group Puts pumps into groups for multiple pump rotations on one tra
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 137mmmmmParametersP170 Flush Pump Picks the number of the pump relay which triggers the flushin
Page 138 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP173 Flush Duration The length of time for each flush cycle that the flush co
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 139mmmmmParametersP201 mA Output Function Alters the mA output/measurement relationship.P202 mA
Page 140 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP203 mA Output Value / Transducer Displays current mA output value for the Po
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 141mmmmmParametersP210 0/4 mA Output Setpoint Sets the process level corresponding to the 0 or
Page 142 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP213 mA Output Max Limit Sets the maximum mA output value (in mA) to be produ
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 143mmmmmParametersmA Output Failsafe (P219) P219 mA Output Failsafe Use for failsafe operation,
Page 144 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP251 0 or 4 mA Input Level Shows the process level corresponding to the 0 or
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 145mmmmmParametersP254 Scaled mA Input Value Shows the resulting level value after scaling.This
Page 2 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmHydroRanger 200Manual SymbolsPlease note their use carefully.Configuration ExamplesThe co
Page 146 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP275 Scaled Discrete Input ValueShows the current value of the discrete input
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 147mmmmmParametersP300 Temperature, Transducer MaximumShows the highest temperature encountered
Page 148 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP303 Temperature, Sensor MinimumShows the lowest temperature encountered, as
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 149mmmmmParametersPump Records (P310 to P312)These features to identify pump usage and if the a
Page 150 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP312 Pump Run Ons View or reset the accumulated number of times the displayed
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 151mmmmmParametersLCD Totalizer (P322 and P323) Use these features to view, reset, or preset th
Page 152 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersProfile Records (P330 to P337) These features can record up to ten Echo profi
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 153mmmmmParametersRecording a profile manuallyPress TRANSDUCER to fire the transducer and rec
Page 154 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP332 Auto Record TransducerSpecifies the Transducer Point Number for which Au
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 155mmmmmParametersP334 Auto Record ON SetpointEnter the critical level which, in conjunction wi
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 3mmmmmSpecificationsSpecificationsPowerAC version• 100-230 Vac ± 15%, 50 / 60 Hz, 36 VA (17W)1•
Page 156 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP337 Auto Record LOE TimeLimits Auto Profile Records from being saved unless
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 157mmmmmParametersP342 Start UpsThe number of times power has been applied since the Date Of Ma
Page 158 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersIf your PMD is not listed, select the appropriate Universal Flow Calculation.
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 159mmmmmParametersExample ExponentsP602 Primary Measuring Device Dimensions The dimensions of t
Page 160 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP603 Maximum Head The level value associated with Maximum Flow, in Units (P00
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 161mmmmmParametersP605 Zero Head The distance above Empty (P006) in Units (P005) representing z
Page 162 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP607 Flowrate Decimal The maximum number of decimal places to be displayed.In
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 163mmmmmParametersThis parameter is enabled only if the primary measuring device (PMD) supports
Page 164 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP620 Low Flow Cutoff Eliminates totalizer activity for flows at or below the
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 165mmmmmParametersPumped Volume Totalizer (P622) The Pumped Volume Totalizer is a HydroRanger 2
Page 4 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmSpecificationsRange• 0.3m (1 ft) to 15m (50 ft), dependent on transducerAccuracy• 0.25% o
Page 166 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersTotalizer (P630 to P645) P630 LCD Totalized Multiplier Use this feature if th
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 167mmmmmParametersP633 LCD Totalized Decimal Position Enter the maximum number of decimal place
Page 168 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP641 Flow Sampler Mantissa Use this feature in conjunction with Flow Sampler
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 169mmmmmParametersP645 Relay Duration Use this feature (if desired) to adjust the minimum conta
Page 170 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersBefore using this feature, verify the following parameters are correct:•Empty
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 171mmmmmParametersP652 Offset CorrectionThe value altered when an Offset Calibration is perform
Page 172 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersTemperature Compensation (P660 to P664)P660 Temp SourceSource of the temperat
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 173mmmmmParametersP663 Temperature Transducer AllocationThis feature may only be used for diffe
Page 174 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersRate (P700 to P708) These parameters determine how material level changes are
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 175mmmmmParametersP702 Filling IndicatorThe fill rate required to activate the LCD Filling indi
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 5mmmmmSpecificationsFixed• 0.17% per °C deviation from programmed valueOutputsTransducer drive•
Page 176 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersEnter the time or distance interval over which the Rate Value is to be calcul
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 177mmmmmParametersP708 Volume Rate Display The rate of change of volume in percent of maximum v
Page 178 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersIf a material agitator (mixer) is used in the vessel monitored, set Echo Lock
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 179mmmmmParametersP713 Echo Lock WindowAdjusts the size of the Echo Lock Window.The Echo Lock W
Page 180 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersThis feature may only be used to adjust the delay before the next point is sc
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 181mmmmmParametersSelecting the Auxiliary Reading operation1. Press READING to display the Au
Page 182 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP733 Scroll AccessUse this feature to select the parameter scroll access opti
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 183mmmmmParametersSmartLinx Reserved (750 to 769) These parameters are reserved for optional Sm
Page 184 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP772 Baud RateThe communication rate with the master device.This specifies th
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 185mmmmmParametersP775 Stop BitsThe number of bits between the data bits.P778 Modem AvailableSe
© Siemens Milltronics Process Instruments Inc. 2002Safety GuidelinesWarning notices must be observed to ensure personal safety as well as that of oth
Page 6 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmSpecificationsDiscrete (2)• 10-50Vdc switching level• logical 0 = < 0.5 Vdc• logical 1
Page 186 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP782 Parameter Index LocationDetermines where index information is stored for
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 187mmmmmParametersP791 Bus ErrorIndicates if an error condition is occurring on the bus.P792 Bu
Page 188 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP801 Range ExtensionAllows the material level to fall below the Empty setting
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 189mmmmmParametersP803 Shot / Pulse ModeDetermines what type of ultrasonic shots are fired.Use
Page 190 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP805 Echo ConfidenceDisplays the echo confidence of the measurement echo from
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 191mmmmmParametersAdvanced Echo Processing (P810 to P825) The following parameters are for auth
Page 192 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersAny combination of the following Scope Displays is available. Two methods to
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 193mmmmmParametersP816 Echo Time RawThe time (in ms) from the transmit pulse to the processed e
Page 194 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP818 Profile Pointer DistanceThe distance between the transducer face and the
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 195mmmmmParametersUse this to select the algorithm(s) the Sonic Intelligence®echo selection is
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 7mmmmmInstallationInstallationMountingMounting LocationsRecommended• Ambient temperature is alw
Page 196 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP823 Reform Echo Smoothes jagged peaks in the echo profile.Use this feature,
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 197mmmmmParametersAdvanced TVT Adjustment (P830 to P835) The following parameters are for autho
Page 198 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP832 TVT Shaper AdjustAllows manual adjustment of the TVT curve.Use this feat
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 199mmmmmParametersP833 TVT Start MinUse this feature to adjust the TVT Curve height to ignore f
Page 200 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersAdvanced Shot Adjustment (P840 to P852) These parameters are for Siemens Mill
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 201mmmmmParametersP843 Long Shot FrequencyAdjust the long shot transmit pulse frequency (in kHz
Page 202 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP850 Short Shot BiasUse this feature to slant the echo evaluation in favour o
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 203mmmmmParametersTest (P900 to P913)Test Parameters are intended for use by Siemens Milltronic
Page 204 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP904 KeypadPress ENTER , then press each keypad key in the following sequen
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 205mmmmmParametersP910 Toggle RelaysUsed to energize and de-energize relays directly.Enter the
Page 8 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmInstallationMounting InstructionsMount the unit where there is sufficient room to swing l
Page 206 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP913 Sensor TemperatureAccess this parameter to display the temperature in °C
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 207mmmmmParametersReading Measurements by OperationP921 Material MeasurementThe distance in Uni
Page 208 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersP924 Volume Measurement The calculated vessel capacity in Max Volume (P051) o
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 209mmmmmParametersMaster Reset (P999) This feature resets all parameters to original values.Use
Page 210 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmParametersNotes
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 211mmmmmAppendix A: Index TypesGeneral Appendix A: Index typesIndex typesName Description # of
Page 212 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmAppendix B: Tech ReferenceGeneral Appendix B: Technical ReferenceTransmit PulseThe tran
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 213mmmmmAppendix B: Tech ReferenceDolphin Plus DisplayDistance CalculationTo calculate the tran
Page 214 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmAppendix B: Tech ReferenceThe HydroRanger 200 automatic temperature compensation is bas
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 215mmmmmAppendix B: Tech ReferenceEnter a Level Breakpoint at each point where the level/volume
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 9mmmmmInstallationCable routed through a conduit:1. Remove the four mounting screws holding the
Page 216 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmAppendix B: Tech ReferenceEnter a Head Breakpoint at each point where the head/flow cur
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 217mmmmmAppendix B: Tech ReferenceMaximum Process SpeedThe HydroRanger 200’s ability to respond
Page 218 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmAppendix C: TroubleshootingGeneral Appendix C: TroubleshootingIf the setup becomes too
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 219mmmmmAppendix C: TroubleshootingNoise ProblemsIncorrect readings can be the result of noise
Page 220 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmAppendix C: TroubleshootingDetermine the Noise SourceDisconnect the transducer from the
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 221mmmmmAppendix C: TroubleshootingFiltering cables is an option but is not recommended unless
Page 222 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmAppendix C: TroubleshootingMeasurement DifficultiesIf the Failsafe Timer (P070) expires
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 223mmmmmAppendix C: TroubleshootingDisplaying Echo Confidence in the RUN modePress and hold f
Page 224 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmAppendix C: TroubleshootingIf an obstruction cannot be removed or avoided, adjust the T
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 225mmmmmAppendix C: TroubleshootingWrong ReadingIf the Reading is erratic, or jumps to some inc
Page 10 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmInstallationInstalling the BatteryThe battery (Rayovac BR2032) has a ten-year life expec
Page 226 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmAppendix C: TroubleshootingNormal Ring DownPoor Ring DownRing down times that extend pa
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 227mmmmmAppendix D: Pump Control General Appendix D: Pump Control ReferenceThe HydroRanger 200
Page 228 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmAppendix D: Pump Control AlternateStarts pumps based on the duty schedule and always le
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 229mmmmmAppendix D: Pump Control Relay TableThe following table shows relay status when each se
Page 230 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmAppendix D: Pump Control Alternate Duty Backup (P111 = 53) Alternates the lead pump eac
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 231mmmmmAppendix D: Pump Control Service Ratio Duty Assist (P111 = 54) Selects the lead pump ba
Page 232 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmAppendix D: Pump Control First In First Out (FIFO) (P111 = 56) Selects the lead pump ba
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 233Appendix E: Updfate SoftwareGeneral Appendix E: Updating SoftwareUpdating SoftwareTo update
Page 234 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmGeneral Appendix FGeneral Appendix F: UpgradingThe following procedure will assist you
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 235mmmmmGeneral Appendix FConnecting a transducer with RG62 coaxial extension cableIf you are r
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 11mmmmmWiringWiringPlease note:• Verify that all system components are installed in accordance
Page 236 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmGeneral Appendix FHydroRanger to HydroRanger 200 ParametersThe HydroRanger 200 uses the
7ML1998IFC01 HydroRanger 200 – INSTRUCTION MANUAL Page 237mmmmmIndexIndexAaccuracy 4adding new functions 233alarm 47bounded 49cable fault 49common pa
Page 238 HydroRanger 200 – INSTRUCTION MANUAL 7ML1998IFC01mmmmmIndexcut throat 75H flume 71Leopold Lagco 74Palmer Bowlus 70Parshall 73, 77rectangular
7ML1998IFC01 HydroRanger 200 – INSTRUCTION MANUAL Page 239mmmmmIndexoverview 95reading 96writing 98parameter indexing 29, 96parametersadvanced echo p
Page 240 HydroRanger 200 – INSTRUCTION MANUAL 7ML1998IFC01mmmmmIndexrecording profiles 153register map 91discrete inputs 94input/output 93mA output 94
7ML1998IFC01 HydroRanger 200 – INSTRUCTION MANUAL Page 241mmmmmIndexcalculation 214characterization chart 45dimensions 33readings 44tank shape 44univ
*7ML19981Fc01*Rev. 1.2
Page 12 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmWiringTerminal BoardThe terminal board on the HydroRanger 200 allows all inputs and outp
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 13mmmmmWiringTransducersA 0.1 µF (100V or greater) capacitor is included with the HydroRanger 2
Page 14 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmWiringTemperature SensorAccurate temperature readings are critical to accurate level mea
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 15mmmmmWiringLevel System SynchronizationWhen using multiple ultrasonic level monitors, be sure
immmmmTable of ContentsTable of ContentsHydroRanger 200 ...
Page 16 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmWiringDigital CommunicationsWiring the HydroRanger 200 for communications allows it to b
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 17mmmmmOperationOperating the HydroRanger 200The HydroRanger has two modes of operation: RUN an
Page 18 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmOperationIcons indicating index type (Item 1) edited in PROGRAM mode:Readings in RUN Mod
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 19mmmmmOperationStatus ParametersStatus parameters give the operating status of the HydroRanger
Page 20 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmOperationExampleTo reference the displayed level to sea level, enter the distance in Uni
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 21mmmmmOperationSetting a Specific Auxiliary ReadingTo display a second auxiliary reading, pres
Page 22 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmOperationStarting PROGRAM ModeHand ProgrammerThe hand programmer gives you direct access
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 23mmmmmOperationProgrammer KeysKeys Programming Mode Run Mode1 8-digit Totalizer (toggle) 2 Pum
Page 24 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmOperationDolphin PlusUse Dolphin Plus software to configure, monitor, tune, and diagnose
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 25mmmmmOperationDolphin Plus Toolbar ButtonsThe toolbar buttons provide quick access to Dolphin
iimmmmmTable of ContentsUsing Units or Percent (%) ...
Page 26 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmOperationActivating the HydroRanger 200All the instructions in the following procedures
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 27mmmmmOperationWhen P000 is set to 1954, programming is enabled. To disable programming, enter
Page 28 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmOperationParameter ResetReturning a parameter to factory default.1. Display the appropri
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 29mmmmmOperationChanging Parameters (Dolphin Plus)The other method for changing parameter value
Page 30 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmOperationHydroRanger 200 DisplayThe index number and the index values are displayed abov
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 31mmmmmOperationPrimary and Secondary IndexesPrimary Index: relates to direct input or output a
Page 32 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmOperationExample P054 provides up to 32 breakpoint levels used with P055 (Volume Breakpo
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 33mmmmmOperationMeasurement ConditionsThe following information will help you configure your Hy
Page 34 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmRelays3RelaysRelays are the primary controls of external devices such as pumps or alarms
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 35mmmmmRelaysTemperature In high alarm, the function goes on when the temperature rises to the
iiimmmmmTable of ContentsOther Pump Control Algorithms ...
Page 36 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmRelaysRelay Status – Non Run Modes When the fail-safe timer expires, pump control relays
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 37mmmmmRelaysP111–Control FunctionsThe control function energizes the relay coils. When the ins
Page 38 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmRelaysFunctions affected by setpoint are configured by parameters that determine the app
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 39mmmmmRelaysReservoir 13Pump up with the following level settings:ParameterRelay #123456P111 5
Page 40 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmBackup Level OverrideBackup Level OverrideBackup level override provides the option of o
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 41mmmmmDiscrete InputsDiscrete InputsWiring the Discrete InputsNormal state is standard operati
Page 42 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmmA I/OmA I/OTo integrate the HydroRanger 200 with other equipment, use the mA input and
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 43mmmmmmA I/OCalibrating 4 mA Output1. Connect the mA receiving device to the HydroRanger 200.
Page 44 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmVolumeVolume Volume is used in two situations:1. Calculate and display volume instead of
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 45mmmmmVolumeCharacterization Chart If you cannot use a pre-defined tank, then use one of the u
ivmmmmmTable of ContentsSimulating a Level Cycle ...
Page 46 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmVolumeParameter Transducer Index Value DescriptionP054 110.0Determines the Level breakp
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 47mmmmmAlarmsAlarmsLevelThe level alarm is the most common. Use this alarm to warn you when you
Page 48 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmAlarmsSetting Simple Level AlarmsTo set Relay Five to a standard level alarm (Hi Hi, Hi,
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 49mmmmmAlarmsIn Bounds / Out of Bounds Range Use the bounded range alarms to detect when the le
Page 50 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmAlarmsWith P112 and P113, you can set a high alarm (P112 > P113) or a low alarm (P112
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 51mmmmmPump ControlPump ControlSetting a Pump Down GroupExample: Sewage Wet WellSetting a group
Page 52 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmPump ControlSet Relays to ALTERNATE DUTY ASSIST Set the ON SetpointsSet the OFF Setpoint
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 53mmmmmPump ControlSet the Common Parameters Prerequisite: Substitute the details of your appli
Page 54 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmPump ControlOther Pump Control AlgorithmsSet Relays to ALTERNATE DUTY BACKUP Set the Rel
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 55mmmmmPump ControlSet Relays to FIXED DUTY BACKUP Set the Relay ON Setpoints Set the Relay OFF
vmmmmmTable of ContentsBit Values ...
Page 56 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmPump ControlSet Relays to FIRST IN FIRST OUT (FIFO) ASSIST Set the Relay ON Setpoints Se
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 57mmmmmPump ControlWhen the first ON setpoint is reached, the pumps will start, one by one, unt
Page 58 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmPump ControlTotalizing Pumped Volume Prerequisite: the volume of the vessel must be know
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 59mmmmmPump ControlSetting a Pump to Run On When you need to pump below the normal OFF setpoint
Page 60 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmPump ControlExample :A range of 0.5 meters is used to vary the setpoint. The randomly-se
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 61mmmmmPump ControlExample: The flush valve connects to Relay Four and the watched pump is on R
Page 62 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01Rake (Screen) Control Rake (Screen) Control Screens or rakes are mounted on the inflow channe
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 63mmmmmRake (Screen) ControlSetting the Common ParametersPrerequisite: Substitute the details o
Page 64 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmExternal Totalizers and Flow Samplers External Totalizers and Flow Samplers External tot
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 65mmmmmExternal Totalizers and Flow Samplers The source of units depends on the operation:Flow
vimmmmmTable of ContentsRelays (P100 to P119) ...
Page 66 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmOpen Channel MonitoringOpen Channel Monitoring (OCM) An OCM installation is defined one
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 67mmmmmOpen Channel MonitoringSetting Zero HeadMany PMDs start flowing higher than the traditio
Page 68 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmOpen Channel MonitoringSetting Totalized VolumeTo display the totalized volume on the LC
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 69mmmmmOpen Channel MonitoringBS-3680 / ISO 4359 Rectangular FlumeParameter Index ValueP600 G 6
Page 70 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmOpen Channel MonitoringPalmer Bowlus FlumeApplication Information• Sized by pipe diamete
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 71mmmmmOpen Channel MonitoringH Flume• Sized by maximum depth of flume• Approach is preferably
Page 72 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmOpen Channel MonitoringPMDs with Exponential Flow to Head FunctionFor Primary Measuring
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 73mmmmmOpen Channel MonitoringNon-Applicable Weir ProfilesFlows through these weirs can be meas
Page 74 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmOpen Channel MonitoringLeopold Lagco FlumeApplication Information• Designed to be instal
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 75mmmmmOpen Channel MonitoringCut Throat FlumeApplication Information• Similar to Parshall flum
viimmmmmTable of ContentsP270 Discrete Input Function...
Page 76 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmOpen Channel MonitoringUniversal Calculation SupportWhen the primary measuring device (P
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 77mmmmmOpen Channel MonitoringExample FlumesThese example flumes would both require a universal
Page 78 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmOpen Channel MonitoringNotes
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 79mmmmmTesting the ConfigurationTesting the ConfigurationAfter programming the unit, you must t
Page 80 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmTesting the ConfigurationThe effect of the ARROW key is determined by the state (rate of
7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 81mmmmmTesting the ConfigurationI/O CheckoutAfter you have installed the unit, we recommend tha
Page 82 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01mmmmmTesting the ConfigurationMonitor system performance carefully, under all anticipated ope
7ML19981FCO1 HydroRanger 200 – INSTRUCTION MANUAL Page 83mmmmmCommunicationsHydroRanger 200 CommunicationsHydroRanger 200 Communication SystemsThe Hyd
Page 84 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FCO1mmmmmCommunicationsCommunication SystemsThe HydroRanger 200 is capable of communicating with
7ML19981FCO1 HydroRanger 200 – INSTRUCTION MANUAL Page 85mmmmmCommunicationsSmartLinxOther protocols are available through optional SmartLinx communic