Page 214 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01
mmmmm
Appendix B: Tech Reference
The HydroRanger 200 automatic temperature compensation is based on the sound
velocity / temperature characteristics of “air” and may not be suitable for the atmosphere
present. If the atmosphere temperature is variable, perform frequent Sound Velocity
Calibrations to optimize measurement accuracy.
Sound Velocity calibration frequency may be determined with experience. If the sound
velocity in two or more vessels is always similar, future calibrations may be performed on
one vessel and the resultant Velocity (P653) entered directly for the other vessel(s).
If the sound velocity of a vessel atmosphere is found to be repeatable at specific
temperatures, a chart or curve may be developed. Then, rather than performing a Sound
Velocity Calibration each time the vessel temperature changes significantly, the
anticipated Velocity (P653) may be entered directly.
Scanning
When echo processing is complete, (if more than 1 vessel is monitored) the scanning
relay changes state to supply the transmit pulse to the other transducer after the Scan
Delay (P727).
Scan Delay is automatically set by Maximum Process Speed (P003). When high speed
scanning is required (sometimes the case for equipment position monitoring), the Scan
Delay may be reduced. Reduce the Scan Delay only as required, otherwise premature
scanning relay fatigue could occur.
When a single point HydroRanger 200 is programmed for differential or average level
Operation (P001 = 4 or 5), two transducers of the same type must be used.
Volume Calculation
The unit provides a variety of volume calculation features (P050 to P055).
If the vessel does not match any of the eight preset Tank Shape calculations, a Universal
Volume calculation may be used. Use the level/volume graph or chart provided by the
vessel fabricator (or create one based on the vessel dimensions). Based on the graph,
choose the Universal Volume calculation, and select the level vs. volume breakpoints to
be entered (32 max). Generally, the more breakpoints entered, the greater the accuracy.
Universal, Linear (P050 = 9)
This volume calculation creates a piece-wise linear approximation of the level/volume
curve. This option provides best results if the curve has sharp angles joining relatively
linear sections.