7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 173
mmmmm
Parameters
P663 Temperature Transducer Allocation
This feature may only be used for differential or average Operation (P001 = 4 or 5).
As preset, the temperature measurements of Ultrasonic / Temperature Transducer One
and Two are allocated to Points 1 and 2 respectively.
Use this feature if the temperature measurement from both transducers should be
identical, but one is located close to a radiant heat source. Allocate the temperature
measurement of the other transducer to both transducer Point Numbers.
Enter the number of the Transducer whose temperature measurement will be used for
the distance calculation of the Point Number displayed. When both transducers are
allocated to a Point Number, the temperature measurements from each are averaged.
P664 Temperature
View the transducer temperature in °C.
Value is displayed when is pressed in RUN mode (see
Readings in Run Mode
on
page 18).
If Temp Source (P660) is set to any value other than Fixed Temp, the value displayed is the
temperature measured. If Temp Source is set to Fixed Temp, the P661 value is displayed.
Primary Index Transducer
Values
1*Transducer One
2Transducer Two
1:2 Transducer One and Two average
Related
P651 Sound Velocity Calibration
P653 Velocity
P654 Velocity at 20°C
Primary Index Transducer
Values Range: -50 to 150 (view only)
Altered By P660 Temp Source
Related
P651 Sound Velocity Calibration
P653 Velocity
P654 Velocity at 20°C
P661 Temp Fixed