7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 191
mmmmm
Parameters
Advanced Echo Processing (P810 to P825)
The following parameters are for authorized Siemens Milltronics Service personnel or
technicians familiar with Siemens Milltronics echo processing techniques.
Anatomy of an Echo Profile
The relevant parts of an echo profile are listed here. These are visible in either Dolphin
Plus or an oscilloscope.
P810 Scope Displays
Captures echo profiles for display on an oscilloscope.
Use this feature to monitor the effects of Echo Processing changes.
Connect an oscilloscope to Mother Board TP2 (Scope), TP5 (DCOM), and TP6 (Sync).
Sweep = 10 us / div. to 1 ms / div. (x 100 for real time)
Amplitude = 1 V / div.
Trigger = external
Primary Index Transducer
Values
Display: P, C, n, u, _
Preset: _ _ _ _ (display is off)
Related P832 TVT Shaper Adjust
Echo lock window
Echo marker
Selected echo
Echo profile
TVT curve
Note: Use of an analog oscilloscope is preferred over a digital oscilloscope.