7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 3
mmmmm
Specifications
Specifications
Power
AC version
• 100-230 Vac ± 15%, 50 / 60 Hz, 36 VA (17W)
1
• fuse: F3: 2 AG, Slow Blow, 0.375A, 250V
DC version
• 12-30 Vdc, 20W
1
• fuse: F3: 2 AG, Slow Blow, 2A, 250V
Transmitter fuse
• F1: Belling Lee, L754, 4000A HRC, ceramic type, 100mA, 250V
Temperature Sensor fuse
• F2: Belling Lee, L754, 4000A HRC, ceramic type, 50mA, 250V
Mounting
Location
• indoor / outdoor
Altitude
• 2000m max.
Ambient temperature
• -20 to 50°C (-5 to 122°F)
Relative humidity
• suitable for outdoors (Type / Nema 4x, IP65 Enclosure)
Installation category
•II
Pollution degree
•4
1.
Power consumption is listed at maximum.