Page 148 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01
mmmmm
Parameters
P303 Temperature, Sensor Minimum
Shows the lowest temperature encountered, as measured by the TS-3 Temperature
Sensor (if applicable).
Press CLEAR keys to reset the log after an open circuit on the transducer wiring.
Record Readings (P304 and P305)
This identifies the occurrence of the record high and low level readings. Press CLEAR
keys to reset these values once the installation is working correctly.
P304 Reading Max
Shows the highest Reading calculated (in normal Reading units or %).
P305 Reading Min
Shows the lowest Reading calculated (in normal Reading units or %).
Primary Index Global
Values
Range: - 50 to 150°C (view only)
Preset: 150°C
Related • P302 Temperature, Sensor Max
Primary Index Level
Values Range: -999 to 9999 (view only)
Related • P305 Reading Min
Primary Index Level
Values Range: -999 to 9999 (view only)
Related • P304 Reading Max