System Overview / Operating Elements
Operating Elements and Displays in the Examination Room
58 / 68 AXA4-100.620.08.01.02 Operator Manual
Table position
o Angle of lateral table tilt
o Angle of the (reverse) Trendelenburg table tilt
o Angle of table swivel about vertical axis Artis dTC/dTA
o Longitudinal position of the tabletop
"0" is displayed if the tabletop is in the end position at the foot end.
o Transverse position of the tabletop
"0" is displayed if the center of the tabletop coincides with the beam center.
o Isocenter-tabletop distance (T.O.D.)
The data refer to the current patient position relative to the patient.
In Artis dMP, the position of the C-arm also influences the coordinates of the
table. Example: (Artis dMP with right mounting)
Tabletop to the right has the same effect as C-arm to the left, with Head First
- Supine the FD will move toward the head of the patient, the longitudinal po-
sition of the tabletop increases in the positive direction.
➩ “Patient coordinates” on page 61
The data refers to the top edge of the tabletop.
The mattress must be taken into account additionally.