4 / 40 AXA4-100.620.08.01.02 Operator Manual
Example of exam set, view with roadmap programs
The fluoroscopy/roadmap programs are always available in three versions, which
usually differ with regard to their dose.
Current exam set
You can view and edit the acquisition/fluoroscopy/roadmap programs which are
stored in the database and arranged in exam sets.
Every acquisition, fluoroscopy, or roadmap program is stored once in the system
and can be used in one or more exam sets.
Acquisition programs
Roadmap programs
The first acquisition program should always be a DR or DSA program when per-
forming a DR-Scanning, PERISTEPPING, PERIVISION, DR-DYNAVISION or DSA-
DYNAVISION acquisition.
The exam sets on your system depend on the type of system and the profile set.
A number of exam sets are preinstalled on the system when it is delivered.
u Ask your application specialist for further information about preinstalled exam
sets.
The settings of the current programs do not have to be the same as the settings
of the programs stored in the database.
The values shown in the images are only explanatory. They have no clinical
meaning.