System Operation
Unit Movements
24 / 94 AXA4-100.620.08.01.02 Operator Manual
Table movements
Raising / lowering the table
u Press the key for the required movement and keep it pressed until the re-
quired table height is reached.
The table height is changed immediately.
The speed of the upward table movement depends on the pressure exerted on
the key:
o light pressure: low speed
o strong pressure: high speed
Stopping in the isocenter
ISO stop You can program the table to stop automatically when the isocenter is reached.
u Set the table height for the isocenter.
u Press the Isocenter key.
The LED lights up.
The current table height is saved as isocenter height.
The table stops every time this position is reached.
The displays of the table coordinates X-Position, Y-Position, and IO-
Height are set to zero and show the relative coordinates (relative to the
patient with reference to the patient position) as long as the LED is lit.
No stop u Press the Isocenter key again.
The LED goes out.
The table moves continuously through the isocenter.
The displays of the table coordinates X-Position, Y-Position, and IO-
Height indicate absolute (table) coordinates again.
The ISO stop function is automatically deactivated when a new patient is regis-
tered.
In the case of Artis dMP systems, downward table movements can result in a
collision with the C-arm. In that case, the movement stops.
If you press the key again, the table will no longer move downward, but (depend-
ing on configuration) the stand will move downward!