Examination
Preparing the Examination - Fluoroscopy - Acquisition
AXIOM Artis AXA4-100.620.08.01.02 43 / 64
Blocking radiation
In some cases, for instance when the patient is repositioned or while cleaning, it
can be useful to block radiation to prevent it from being released inadvertently.
At the touchscreen control (only)
u Activate Block Radiation on the Exam task card.
– The selection field appears selected.
– It is not possible to release radiation.
Fluoroscopy
Single plane Using the FLUORO pedals
u Operate the FLUORO pedal of the footswitch.
Biplane With biplane systems you can perform fluoroscopy in one plane or in both planes:
u Operate the correct FLUORO pedal of the footswitch.
– The radiation indicator(s) light(s) up during radiation.
– Fluoroscopy images appear on the monitor.
– In each case the last fluoroscopy image remains displayed on the monitor
after fluoroscopy is completed (LIH = Last Image Hold).
A&B
A
B