Troubleshooting
System Messages / Troubleshooting
AXIOM Artis AXA4-100.620.08.01.02 11 / 32
Resetting the unit computer
If all these measures do not help, then there could be an error in the unit comput-
er (SCU). In order to restore a defined status, you can reset the computer:
u Simultaneously press the buttons I and II as well as the emergency STOP
button on the stand/C-arm control module (SCM).
The unit computer is restarted and should be ready for operation after a
short time.
u Do not forget to pull the emergency STOP button again .
Restart necessary!
Some faults make it necessary to restart the imaging system or the entire sys-
tem. For most faults, it is enough to restart only the imaging system.
Message o "... Restart"
o "... Restart the imaging system"
Cause The system is in a state that makes a restart necessary.
Measure u Close the imaging system application and restart.
“Restarting after a fault” on page 18
Please pay attention to “Bypass fluoroscopy (emergency operation)” on
page 14
u Register the patient again.
u Continue with the examination.