Image Viewing and Image Processing
DSA Postprocessing
82 / 112 AXA4-100.620.08.01.02 Operator Manual
Scope of action Each pixelshift action (automatic or manual) always applies to the current image
and all images up to the end of the scene.
It is therefore advisable to start with pixelshift at the beginning of the scene. You
can then scroll through the scene to the end. Each time motion artifacts appear
(if the patient or unit has moved), you can make a further pixelshift correction.
Starting pixelshift
At the system console
u Click on this icon.
– A window appears.
Movement Movement
Movement
Pixelshift
action
Pixelshift
action
Pixelshift
action
Scene
The pixelshift magnitudes in the horizontal (x) and vertical (y) direction are dis-
played in the Pixelshift window.