Accessories and Auxiliary Devices
Accessories for Artis (d)MP
AXIOM Artis AXA4-100.620.08.01.02 37 / 54
Compression unit
Use You can use the compression unit both for compressing disturbing tissue layers
and for immobilizing restless patients.
AL equivalent ≤ 0.8 mm
Attaching the compression unit
u Insert the tensioning part (A) of the compression unit into the tabletop.
u Secure the tensioning part with the hand screw (1).
u Release the belt lock by pulling the lever (2) toward you.
– You can now unroll the compression belt.
u Insert the guide part (B) into the opposite side of the tabletop and secure it
with the hand screw (3).
u Pump the lever (2) to tighten the compression belt.
u Pull the lever (2) toward you to release the compression belt.
2
3
1
A
B