System Overview / Operating Elements
Operating Elements and Displays in the Examination Room
AXIOM Artis AXA4-100.620.08.01.02 47 / 68
Cable with plug
On newer systems, the footswitch cable is equipped with a plug.
This symbol marks the appropriate plug for the footswitch at the base of the pa-
tient table.
Signs and indicators on the wireless footswitch
The wireless footswitches are marked with a colored sign of different shape to
avoid a mix up for installations with more than one wireless footswitch.
The same marks are attached to the respective table and room during installation.
Status LEDs There are 3 LEDs showing the status of the footswitch.
o Activity LED
LED lights: Footswitch is on and is in active mode.
LED off: Footswitch is inactive.
– It is too far away from the table
– The Artis system is switched off
– The batteries are completely discharged
– The Footswitch is in “sleep mode”. Press a pedal swiftly to activate it
again.
o Transmission LED lights: Any footswitch pedal is being pressed.
On Artis dMP:
If the safety bracket is activated, all movements stop and the LED turns on.
o Battery LED
– LED off: Battery is sufficiently charged, footswitch is not connected to the
power adapter.
– LED flashing yellow: Battery needs charging, remaining capacity is < 4h.
– LED lights yellow: Footswitch is connected to the power adapter, batteries
are being charged.
– LED lights green: Footswitch connected to the power adapter, batteries
are fully charged.
This symbol marks an RF transmitting component.