Accessories and Auxiliary Devices
Miscellaneous Accessories
AXIOM Artis AXA4-100.620.08.01.02 45 / 54
Technical data o The lead equivalent value of all lead rubber parts is 0.5 mm (according to
EN 61331-3:1999).
o The heat resistance is approximately 80°C.
Cleaning and
disinfection
Slightly soiled lead leaves and metal parts can be cleaned with a soft cloth and
lukewarm water with a household detergent.
Heavy dirt and grime should be wiped off with a linen cloth moistened with alco-
hol and then rinsed with clear water.
Remove blood spots best of all with cold water.
Remove contrast medium spots best of all with warm water.
Use a disinfectant commercially available for medical use to disinfect the radia-
tion protection device.
Then wipe it off with clear water.
The lower body radiation protection is not suitable for supporting loads! This
means, it cannot bear additional weight of, for instance, lead-rubber aprons.
This could pull down the lower body radiation protection.
u Do not hang any objects on it.
The swiveling part of the lower body radiation protection may only be pushed
onto the rail of the fixed part of the radiation protection attached to the base of
the table.
The lower body radiation protection is not monitored by the collision computer.
u Avoid possible collisions while performing unit movements.
u You must swivel the radiation protection for the lower body away from the
patient table in extreme LAO/cranial angulations to avoid collisions with the
X-ray tube.
If several systems are installed, it might be possible to confuse removable parts
of the lower body radiation protection.
u Use only those parts which belong to this system.