AXIOM Artis AXA4-100.620.08.01.02 43 / 54
Accessories and Auxiliary Devices
Miscellaneous Accessories
Accessories for the primary collimator
Dose measuring device (standard in the EU)
The area dose product is measured with a dose measuring chamber during fluo-
roscopy and acquisition.
Its value is displayed on the system console and on the data display.
Part Examination, Chapter Preparing the Examination - Fluoroscopy - Ac-
quisition, “Monitoring the dose”
The values are also listed separately in the examination data for each scene.
Compensating filter
Blooming in the skull area is reduced in carotid artery angiography with the com-
pensating filter.
Two compensating filters in bowl form are available:
o Compensating filter for adults
o Compensating filter for children
Inserting the filter
u Open the tube cover by pressing the buttons on both sides.
u Turn the compensating filter into the correct working position (that is, into the
correct CRAN/CAUD position in relation to the patient).
u Insert the compensating filter into the lower rails of the primary collimator be-
tween the X-ray tube assembly and the dose measuring chamber.
u Push the compensating filter in until it snaps into place so that it is correctly
centered.
Systems with a dose measuring chamber also calculate and display the skin
dose per scene and register it in the examination data.