Examination
Subtracted Fluoroscopy: Roadmap
AXIOM Artis AXA4-100.620.08.01.02 63 / 64
Roadmap
u Operate the corresponding FLUORO pedal.
Perform roadmap as described under “Single plane roadmap” on page 60.
Resetting roadmap Roadmap can be reset separately for each plane:
Reset A o resets roadmap for plane A;
a new roadmap mask image must be created for plane A.
Reset B o resets roadmap for plane B;
a new roadmap mask image must be created for plane B.
Reset A + B o resets roadmap for both planes;
roadmap can be restarted.
With the footswitch
u Press the Reset A or Reset B pedal.
Plane A or plane B will be reset.
Or
u Briefly press the RDMP ON/OFF pedal.
Plane A and plane B will be reset.
At the touchscreen control
u Activate Reset Rdmp.
Reset A
Reset B