Image Viewing and Image Processing
General
16 / 112 AXA4-100.620.08.01.02 Operator Manual
Moving series/scenes/images accidentally acquired under the
wrong patient
u Call up the Patient Browser.
u Select the series/scenes (images).
Drag & Drop
u Drag the selected data to the target position and release the mouse button.
Rearrange is not possible in the following cases:
o Rearrange to the wrong data level
o Rearrange scenes/images to a series acquired with another patient position
o When images were acquired on a different modality (e.g. CT)
Caution
Assignment of series/scenes/images to the wrong patient
Risk that diagnostic data could not be found
Danger of incorrect diagnoses due to allocation of data to wrong patient
u Move data carefully.
Warning
Merging images with different patient position, patient orientation or laterality
of a patient.
Patient orientation and laterality information may be wrong.
u Do not merge data with different patient position, patient orientation and lat-
erality information.