Examination
Preparing the Examination - Fluoroscopy - Acquisition
10 / 64 AXA4-100.620.08.01.02 Operator Manual
Attaching the ECG
For ECG monitoring of the patient or ECG-triggered fluoroscopy:
u Connect the ECG electrodes and cables in such a way that they do not appear
in the beam path.
u Check the ECG curves on the monitor.
u Check the connection of the trigger signal from the ECG unit to the system if
the ECG unit is not permanently installed.
u Check the equipotential bonding.
Preparing for pressure measurement
1
u Adjust the pressure elements to the atrial plane with the three-way valve.
u Open the three-way valve in front of the pressure element to measure the at-
mospheric pressure (air).
u Perform zero pressure measurement compensation at the hemodynamic
measuring system.
Adjusting the units (stand and table)
u Move the patient table to the required working height.
u Position the tabletop.
u Position the stand and the C-arm as required.
Part System Operation, Chapter Unit Movements.
The ECG cables could become tangled with or be broken by moving parts, e.g.
the X-ray tube, during C-arm angulation.
u Pay attention to the ECG cables during unit movements.
1
Only required for cardiac examinations