System Operation
Unit Movements
AXIOM Artis AXA4-100.620.08.01.02 49 / 94
Storing system positions with shortcut (direct
positions I, II, III)
You can assign the three most frequently used system positions to the direct po-
sitions I, II, III:
u Move the unit to the required position.
u Press the Store key.
The LED on the key lights up.
u Press the required position key.
The current system position is stored under the selected direct position
number. The previously stored position is deleted.
The LED on the Store key goes out.
The system returns to its original state.
Cancellation See above.
Deleting programmed positions
You can also delete the stored system positions.
u Press the Programmed Positions key.
The LED on the key lights up.
The Programmed Positions menu appears on the data display.
u Select the memory position (number) you want to delete by tilting the stand/
C-arm joystick (of plane A) backward/forward.
u Press the Store key for longer than 2 seconds.
The memory position is deleted.
The LED on the Store key goes out.
Positions marked with a lock symbol can only be deleted by Siemens Service.