Siemens ARTIS 2 S Service Manual Page 121

  • Download
  • Add to my manuals
  • Print
  • Page
    / 606
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 120
System Operation
Unit Movements
AXIOM Artis AXA4-100.620.08.01.02 15 / 94
Collision protection
Safety equipment
The systems are equipped with a number of safety mechanisms which help pre-
vent collisions during unit movements:
o Dead man's grip (DMG)
Movements take place only as long as the operating element is actuated.
o Collision computer
A computer constantly calculates all possible collisions between the stand(s)/
C-arm(s) and the patient table.
o Speed reduction
The speed is automatically reduced for movements in the collision area or in
particular system positions and movement is stopped, if necessary.
o Collision sensors / infrared sensors
1
In case of a collision, sensors, e.g. on the FD, automatically stop the move-
ment.
o Collision protection bar on the footswitch
1
o Emergency STOP buttons
Movements are stopped immediately when an emergency STOP buttons is
being pressed.
Part Safety (Volume 1
When movements are blocked, a message appears on the data display.
1
Artis dMP only
If a C-arm is moved into the collision zone with the other C-arm (biplane) or with
the patient table, a warning sound (configurable) is emitted and the movement
is slowed down.
Please note that the movement is not stopped, i.e. collisions can occur!
The safety equipment of the system does not release you as the user from your
obligation to pay attention to the patient and the system during unit move-
ments!
In biplane systems, when one stand is moved, the other stand gets out of the
way (deviation movement) to avoid blocking the movement.