Image Viewing and Image Processing
Image Postprocessing
42 / 112 AXA4-100.620.08.01.02 Operator Manual
Defining window and filter values
Information about windowing
With contrast and brightness windowing you can adapt the image to display the
gray scale region of interest with optimal contrast conditions.
The individual pixels of an X-ray image are stored as integer values. These values
represent a range of image brightness divided into increments much finer than
the human eye can perceive. The monitor can display a specific (smaller) range
of brightness values. The purpose of the windowing function is to define the
range to be displayed and the way it should be displayed.
DR images The displayed window area is defined by a pair of numbers.
The values used for non-subtracted images are the window center = WC and the
window width = WW.
DSA images DSA images displayed after subtraction (difference between fill and mask) can be
windowed in the same way as native images.
For subtracted images, the window brightness = WB and the window contrast =
WC are used.
Principle of windowing (display contrast agent/bones black)