Examination
Preparing the Examination - Fluoroscopy - Acquisition
AXIOM Artis AXA4-100.620.08.01.02 5 / 64
Registering the patient and starting examination
u Click on Exam.
A dialog box is displayed in which you must confirm the patient position.
1
u Click on Confirm if the patient position is correct.
The patient is registered.
The Examination task card is placed in the foreground and you can start the ex-
amination immediately.
When the system is in “BYPASS FLUORO” mode, or during radiation, the pa-
tient cannot be registered for examination.
u In this case, click on Preregister.
1
Does not appear, if only the Cardiac profile is configured.
Warning
Incorrect patient position data
Danger of incorrect diagnosis, e.g. confusion of right and left
The examiner is responsible for the correctness of selecting the application and
the consequences resulting from it.
u Check patient position data in the current image to exclude any errors.
You can enter or change missing patient data at a later time.
Part Image Viewing and Image Processing, Chapter General, “Changing
patient data”