System Operation
Unit Movements
AXIOM Artis AXA4-100.620.08.01.02 51 / 94
Movement to or from a position with swiveled table
(Multispace F only)
To perform movements from one position to another, you may have to move the
floor stand and the patient table manually into another swivel position.
This applies especially when changing the position from Patient Transfer Head
Side/Head Side/Left Side to Right Side, Table Rotated / Left Side, Table Ro-
tated.
“Positions of the floor stand (Artis dFC/dFA and Artis dBC/dBA)” on page 11
Selecting the target position
u Select the required target position.
“Selecting the target position” on page 41.
Swiveling the floor stand manually
u Unlock the floor stand by pressing the unlocking button.
“Manual stand movements” on page 50
u Swivel the stand to the required position and move it back and forth slightly
until it engages.
Pay attention to the messages on the data display.