Accessories and Auxiliary Devices
Accessories for Artis (d)MP
40 / 54 AXA4-100.620.08.01.02 Operator Manual
Foot holder
Use For securing the patient during examinations in the Trendelenburg position with
a relaxed spinal column.
Inspection/mainte-
nance
note
Check that the foot holder is in a proper condition at regular intervals of approxi-
mately two months.
Check especially that the fastening rivets, the supporting straps and closing
straps are firmly attached, that their seams are complete and that the leather is
free of cracks and porous areas.
If the foot holder shows any signs of defects, take it out of service and obtain a
replacement through Siemens Uptime Service.
We recommend using a commercially available transparent impregnating pump
spray for leather for cleaning and care.
Attaching the foot holder
u Position the tabletop horizontally.
u Attach the footboard.
u Make sure that the footboard is attached firmly by pulling and pushing it.
u Position the patient in the prone or supine position with his/her feet resting
against the footboard.
u Place the foot holder closely around the patient's ankles.
u Tighten the straps.