Examination
Subtracted Fluoroscopy: Roadmap
60 / 64 AXA4-100.620.08.01.02 Operator Manual
Performing roadmap
Single plane roadmap
First phase (native) At the touchscreen control
u Activate Roadmap in the Exam task card.
At the system console
u Click on the roadmap icon.
The message "Roadmap reset" appears on the data display.
Fluoroscopy
u Actuate the FLUORO pedal and keep it pressed.
– After fluoroscopy has begun a normal fluoroscopy image is displayed
briefly, then the display automatically switches to a subtracted image.
– In the status messages at the system console and on the data display, the
note "Injection" is displayed.
Comments to this Manuals