AXIOM Artis AXA4-100.620.08.01.02 3 / 40
Exam Sets
The Exam Set Editor
Introduction
The AXIOM Artis provides a number of acquisition, fluoroscopy and roadmap pro-
grams for standard examinations.
The acquisition, fluoroscopy and roadmap programs are arranged in exam sets
and stored in a database. The fluoroscopy/acquisition parameters and dose are
determined in the exam sets.
Exam Sets
An exam set comprises
o 3 fluoroscopy programs and possibly
3 assigned roadmap programs, and
o up to 4 acquisition programs.
Example of exam set, view with fluoroscopy programs
The available exam sets depend on the system type and the profile set.
The system comes preinstalled with a number of exam sets.
u For further information on the preinstalled examination sets ask your applica-
tions specialist.
The programs may only be changed by experienced users.
Acquisition programs
Fluoroscopy programs