Siemens ARTIS 2 S Service Manual Page 399

  • Download
  • Add to my manuals
  • Print
  • Page
    / 606
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 398
Exam Sets
Parameters for Exam Sets
AXIOM Artis AXA4-100.620.08.01.02 23 / 40
Image parameters for DSA acquisitions (example)
User parameters
Processing Mode Exposure control and display mode for observation
o Automatic:
with exposure control (automatic adjustment to object transparency)
o Regulation Stop:
stops exposure regulation after a fixed time (e.g. for left ventricular cardiac ex-
aminations)
o DSA NAT:
non-subtracted display
o DSA SUB:
subtracted display
Gain Correction Radition correction in Siemens exposure points
Contrast Medium Contrast medium used (for DSA acquisition only)
o None
o CO
2
o Iodine
o Gadolinium
o Barium