Siemens SOMATOM Sensation Cardiac Version A60 Operations Instructions Page 594

  • Download
  • Add to my manuals
  • Print
  • Page
    / 600
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 593
594
Children
BodyAngio/BodyAngio042s
Indications:
For abdominal CT Angio studies, e.g. vascular abnor-
malities, aneurysms, etc.
A typical range of 20 cm covered in 12.17/10.22 sec.
* The conversion factor for a 7-year-old child, and a
scan range of 200 mm was used.
BodyAngio 2
nd
reconstr.
kV 80
Effective mAs/
Quality ref. mAs
75
Rotation time 0.5/0.42 sec.
Acquisition 16 x 0.75 mm
Slice collimation 0.75 mm
Slice width 3.0 mm 1.0 mm
Feed/Rotation 9.0 mm
Pitch Factor 0.75
Increment 3.0 mm 0.7 mm
Kernel B30f B20f
CTDI
Vol 2.0 mGy
Effective dose Male: 0.93 mSv*
Female: 1.12 mSv*
Page view 593