HeartView CT
Placement of ECG Electrodes
The correct placement of the ECG electrodes is essen-
tial in order to receive a clear ECG signal with marked
R-Waves. Incorrect placement of the electrodes will
result in an unstable ECG signal which is sensitive to
movements of the patient during the scan.
US Version (AHA standard)
White Electrode:
on the right mid-clavicular line, directly below the clav-
icle
Black Electrode:
on the left mid-clavicular line, 6th or 7th intercostal
space
Red Electrode:
right mid-clavicular line, 6th or 7th intercostal space