CoronaryCTA
CoronaryCTAAdaptSpeedVol
Indications:
This spiral scanning protocol is identical to
CoronaryCTARoutine, except that it uses a rotation
time of 0.5 sec. It is intended to provide dose reserves
for obese patients with sufficiently low heart rates. It
can be also used for patients with a stable heart rate at
70 bpm.
Use this protocol for the 3D images display view.
A typical range of 12 cm covering the entire heart can
be done in 16.15 sec.
Topogram:
AP, 512 mm. Approxi-
mately, from the car-
ina to the apex of the
heart.