318
Neck
NeckRoutine
Indications:
Spiral mode for soft Tissue studies in the cervical
region, e.g. tumors, lymphoma, abscesses etc.
A typical range of 20 cm will be covered in 9.83 sec.
Neck
kV 120
Effective mAs/
Quality ref. mAs
150
Rotation Time 0.75 sec.
Acquisition 16 x 1.5 mm
Slice collimation 1.5 mm
Slice width 5.0 mm
Feed/Rotation 18.0 mm
Pitch Factor 0.75
Increment 5.0 mm
Kernel B31s
CTDI
Vol 10.5 mGy
Effective dose Male: 2.35 mSv
Female: 2.31 mSv
Comments to this Manuals