Workflow Information
Test Bolus using CARE Bolus
You can use the CARE Bolus option as a Test Bolus.
How to do it
1.Insert a Bolus tracking via the right mouse button
submenue prior to the spiral.
2.Insert contrast from the right mouse button context
menu.
Hint: By inserting contrast you are interrupting the
Auto range function,and therefore an automatic
start of the spiral is not possible!
3.Start with the Topogramm.
4.Position the premonitoring scan and the spiral.
5.Perform the premonitoring scan,position and accept
the ROI.
6.Start monitoring scans and a short amount of con-
trast (20 ml/2.5 ml/sec.).
Hint: With starting the spiral the system is switching
to the Trigger subtask card. The trigger line is not
shown at this stage.
7.Now you can read the proper delay from the Trigger
card.
8.Insert the delay in the Routine subtask card and load
the spiral.
9.Start spiral and injector with the full amount of con-
trast.