418
Upper Extremities
ExtrCombi
Indications:
Spiral mode for the combination of bone and soft tis-
sue studies, e.g. masses, trauma, disorders of the joint
etc.
A range of 15 cm will be done in 24.73 sec.
Hint
• For image reconstruction of soft tissue, use kernel
B30s/B31s.
ExtrCombi 2
nd
reconstr.
kV 120
Effective mAs/
Quality ref. mAs
90
Rotation time 1.0 sec.
Acquisition 16 x 0.75 mm
Slice collimation 0.75 mm
Slice width 5.0 mm 1.0 mm
Feed/Rotation 6.6 mm
Pitch Factor 0.55
Increment 5.0 mm 0.7 mm
Kernel B60s B60s
CTDI
Vol
7.0 mGy
Effective dose Male: 0.02 mSv
Female: 0.02 mSv