Page view 319
320
Neck
NeckThinSlice
Indications:
Spiral mode using thin slices for soft tissue studies, e.g.
the functional study of the throat.
A typical range of 18 cm will be covered in 16.5 sec.
NeckThinSlice 2
nd
reconstr.
kV 120
Effective mAs/
Quality ref. mAs
150
Rotation Time 0.75 sec.
Acquisition 16 x 0.75 mm
Slice collimation 0.75 mm
Slice width 5.0 mm 1.0 mm
Feed/Rotation 9.0 mm
Pitch Factor 0.75
Increment 5.0 mm 0.7 mm
Kernel B31s B31s
CTDI
Vol
11.7 mGy
Effective dose Male: 2.17 mSv
Female: 2.27 mSv