Siemens SOMATOM Sensation Cardiac Version A60 Operations Instructions Page 442

  • Download
  • Add to my manuals
  • Print
  • Page
    / 600
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 441
442
Vascular
CarotidAngio
Indications:
CT angiography of the carotid arteries, e.g. carotid
stenosis or occlusion, coarse plaques abnormalities of
the carotids or vertebral arteries, etc.
A range of 20 cm including the aorta arch will be cov-
ered in 8.25 sec.
CarotidAngio 2
nd
reconstr.
kV 120
Effective mAs/
Quality ref. mAs
120
Rotation time 0.5 sec.
Acquisition 16 x 0.75 mm
Slice collimation 0.75 mm
Slice width 5.0 mm 1.0 mm
Feed/Rotation 13.8 mm
Pitch Factor 1.15
Increment 5.0 mm 0.7 mm
Kernel B30f B20f
CTDI
Vol
9.4 mGy
Effective dose Male: 2.02 mSv
Female: 1.89 mSv