Spine
SpineThinSlice
Indications:
Spiral mode for thin slice thoracic and lumbar spine
studies, e.g. prolapse, degenerative changes, trauma,
tumors etc.
A range of 16 cm will be covered in 14.83 sec.
Please notice, if you are not satisfied with the Range
preset, adapt the parameters to your needs and link
them to the series.
Spine
ThinSlice
2
nd
recon.
3
rd
recon.
kV 120
Effective mAs/
Quality ref. mAs
300
Rotation Time 0.75 sec.
Acquisition 16 x
0.75 mm
Slice collimation 0.75 mm
Slice width 3.0 mm 1.0 mm 1.0 mm
Feed/Rotation 9.0 mm
Pitch Factor 0.75
Increment 3.0 mm 0.7 mm 0.7 mm
Kernel B20s B20s B60s
CTDI
Vol
23.4 mGy
Effective dose Male: 7.82 mSv
Female: 9.36 mSv