Respiratory Gating
Scanning Information
The expected benefits in radiation treatment planning,
e.g. for lung and abdominal tumors are:
Information about tumor motion in 3D coordinates
and over time
More accurate tumor shape delineation and there-
fore a more precise RTP (radiation therapy planning)
Potential for sparing of healthy tissue, minimization
of PTV (planned target volume) and less side effects
Potential for dose acceleration and higher cure rate
To minimize motion artifacts, two requirements are
mandatory for a CT system:
Fast gantry rotation to raise the temporal resolution
for artifact free images
Prospective Triggering of image acquisition in a
sequential mode or Retrospective Gating of image
reconstruction in a spiral mode based on the record-
ing of the respiratory curve in order to obtain images
during inspiration and expiration phases.
Scan Parameters
A respiratory-gated lung spiral with cone correction
will be provided. To be able to work also with low
breathing rates, a fixed pitch factor of a min. of 0.1 is
needed. Slice widths and collimation are the same as
for cardio spiral.
In the Specials folder different scan protocols are pre-
defined for different breathing rates.