syngo Calcium Scoring
CaScoreSeq
Indications:
This is a sequential scanning protocol with ECG trigger-
ing for coronary calcium scoring studies.
A scan range is predefined with 15.9 cm.
Topogram:
AP, 512 mm.
From the carina to
the apex of the heart.
* depends on heart rate
CaScoreSeq
kV 120
mAs/
Quality ref. mAs
35
Rotation Time 0.37 sec.
Acquisition 12 x 1.5 mm
Slice collimation 1.5 mm
Slice width 3.0 mm
Feed/Scan 18.0 mm
Kernel B35f
Temp. resolution* up to 185 ms
CTDI
Vol
2.5 mGy
Effective dose Male: 0.71 mSv
Female: 0.97 mSv