C-Spine
Indications:
Spiral mode for cervical spine studies, e.g. prolapse,
degenerative changes, trauma, tumors etc.
A range of 16 cm will be covered in 14.83 sec.
C-Spine 2
nd
recon.
kV 120
Effective mAs/
Quality ref. mAs
330
Rotation Time 0.75 sec.
Acquisition 16 x 0.75 mm
Slice
collimation
0.75 mm
Slice width 2.0 mm 1.0 mm
Feed/Rotation 9.0 mm
Pitch Factor 0.75
Increment 2.0 mm 0.7 mm
Kernel B20s B60s
CTDI
Vol 25.7 mGy
Effective dose Male: 4.78 mSv
Female: 4.9 mSv