Siemens SOMATOM Sensation Cardiac Version A60 Operations Instructions Page 391

  • Download
  • Add to my manuals
  • Print
  • Page
    / 396
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 390
391
Specials
HandCARE positions
Application Procedure
1. Load and scan a spiral protocol of the interested
body region.
2. Scroll through the images to define a target slice.
3. Click on “Same TP” under Table position, in the
routine card. Move the table to the desired table
position.
4. Turn on the light marker on the Gantry to localize
the entry point, and then start patient preparation.
5. Open the “Patient Model Dialog”. Check the
check-box “Append”. Select one of the predefined
“CAREVision” scan protocols under “Specials“
protocols and then click “OK”.
6. Select the “HandCARE” position on the scan card.
7. Click “Load” and the “Cancel/Move” to scan on the
preselected table position.
8. Press the footswitch to start the Fluoroscopy.