Siemens SOMATOM Sensation Cardiac Version A60 Operations Instructions Page 382

  • Download
  • Add to my manuals
  • Print
  • Page
    / 396
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 381
How to do it
Trauma
A scan range of 50 cm will be done in 9.0 sec.
PolyTrauma
Two times 2 combined recon jobs are predefined,
head with neck and thorax with abdomen.
382
Specials
Trauma
kV 120
Effective mAs 140
Rotation time 0.5 sec.
Slice collimation 1.5 mm
Slice width 7.0 mm
Feed/Rotation 28.0 mm
Increment 7.0 mm
Kernel B31f
CTDI
Vol
9.8 mGy
Effective dose Male: 7.2 mSv
Female: 8.9 mSv
Head Neck
kV 120 120
Effective mAs 260 150
Slice collimation 1.5 mm 1.5 mm
Slice width 6.0 mm 5.0 mm
Feed/Rotation 18.0 mm 18.0 mm
Rotation time 0.75 sec. 0.5 sec.
Kernel H31s B31f
Increment 6.0 mm 5.0 mm
CTDI
Vol
49.4 mGy 53.3 mGy
Effective dose Male: 2.0 mSv Male: 1.5 mSv
Female: 2.1 mSv Female: 1.6 mSv