95
Enter “50” “52” “55” “58” “60” to find out the best
display of the LAO view.
As result you will get 5 series with the heartphases:
50%, 52%, 55%, 58%, 60%.
If you want to create your own multiphase recon job
in a cardio scan protocol:
Define either an auto or a manual setting, than
select the “Multiphase settings – Protocol”. Press button
“Apply”, close the window with “OK”. Then you can
save the scan protocol as usual and the selected recon
job will be always reconstructed using a multiphase.
Hints
• To get a constant image quality for the Multiphase
reconstruction we recommend to turn ECG Pulsing
“OFF”.
• If you are not sure which heartphases are the best
one for the Multiphase reconstruction, perform
a preview series first.
• Press “Recon“ button for each 3D reconstruction.