Siemens SOMATOM Sensation Cardiac Version A60 Operations Instructions Page 105

  • Download
  • Add to my manuals
  • Print
  • Page
    / 396
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 104
105
syngo Calcium Scoring
Prerequisites
To achieve standardization of the evaluation results
the image material must meet the following require-
ments:
• Only images of one patient must be loaded for
syngo Calcium Scoring evaluation.
• The scans must have been acquired without gantry
tilt (gantry tilt = 0).
• Only original image material must be used that
has neither been edited with 2D tools in the Viewing
task card e.g., Zoom&Pan nor with 3D tools in 3D
e.g., curved mode.
• Only image data sets with the same slice orientation
must be used.
• Only contiguously scanned and reconstructed
sequence or spiral data sets with identical slice
distances and identical thickness must be evaluated.
• The slices must have been scanned at different table
positions.
• The data capacity of the volume must not exceed
1024 images with a matrix size of 512 x 512 pixels.
• Only studies without contrast medium must be used.
Best Calcium Scoring results are achieved when
special scan protocols for image acquisition have been
used.
For this reason, the SOMATOM CT Scanners offer
several Calcium Scoring scan modes (for example:
“CaScoreSeq“ or “CaScoreRoutine“), which use the
advanced CT HeartView trigger method for scanning.
Page view 104
1 2 ... 100 101 102 103 104 105 106 107 108 109 110 ... 395 396

Comments to this Manuals

No comments