Siemens SOMATOM Sensation Cardiac Version A60 Operations Instructions Page 198

  • Download
  • Add to my manuals
  • Print
  • Page
    / 396
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 197
198
Children
HeadSeq05s
Indications:
Sequential mode for routine head studies for
children, with a 0.5 sec. rotation time, e.g. tumors,
hydrocephalus, hemorrhaging, abnormalities, etc.
Contrast medium IV injection
Start delay exam dependent
Flow rate dependent upon needle size/Access site
Total amount 1 – 2 ml per kg of body weight
* The conversion factor for an 8-week-old, and a scan
range of 90 mm was used.
HeadSeq05s
kV 120
Effective mAs 150
Rotation time 0.5 sec.
Slice collimation 1.5 mm
Slice width 4.5 mm
Feed/Scan 18.0 mm
Kernel C30f
CTDI
Vol
28.8 mGy
Effective dose Male: 1.7 mSv*
Female: 1.7 mSv*