Siemens SOMATOM Sensation Cardiac Version A60 Operations Instructions Page 232

  • Download
  • Add to my manuals
  • Print
  • Page
    / 396
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 231
232
Children
NeonateBody/
NeonateBody042s (optional)
Indications:
Spiral mode for routine neonate body studies, e.g.
tumors, abnormalities, malformations, abscesses, etc.
A typical range of 15 cm covered in 4 sec.
Contrast medium IV injection
Start delay exam dependent
Flow rate dependent upon needle size/Access site
Total amount 1 – 2 ml per kg of body weight
NeonateBody
kV 80
Effective mAs 33
Rotation time 0.5/0.42 sec.**
Slice collimation 1.5 mm
Slice width 6.0 mm
Feed/Rotation 18.0 mm
Increment 6.0 mm
Kernel B30f
CTDI
Vol
0.726 mGy
Effective dose Male: 0.5 mSv*
Female: 0.7 mSv*
* The conversion factor for an 8-week-old, and a scan
range of 100 mm was used.
** optional