
312
Pelvis
SI_Joints
Indications:
Spiral mode for the sacroiliac joints, e.g. evaluation
of joint cavity, masses, trauma, dysplasia, necrosis,
congruence evaluations, orthopedic indications etc.
A typical range of 15 cm will be covered in 5.0 sec.
SI_Joints 2
nd
reconstr.
kV 120
Effective mAs 160
Slice collimation 0.75 mm
Slice width 3.0 mm 1.0 mm
Feed/Rotation 9.0 mm
Rotation time 0.5 sec.
Kernel B60f B60f
Increment 3.0 mm 0.7 mm
CTDI
Vol
12.48 mGy
Effective dose Male: 2.4 mSv
Female: 2.1 mSv
For the 2
nd
reconstruction the Autoload into MPRthick
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MPRthick, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset, adapt the parameters to your needs and link
them to the series.
Comments to this Manuals