Siemens SOMATOM Sensation Cardiac Version A60 Operations Instructions Page 296

  • Download
  • Add to my manuals
  • Print
  • Page
    / 396
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 295
296
Abdomen
AbdomenCombi
Indications:
Thin slice scan for soft tissue spiral studies.
A range of 20 cm for liver, pancreas or kidneys will be
covered in 11.0 sec.
AbdCombi 2
nd
reconstr.
kV 120
Effective mAs 160
Slice collimation 0.75 mm
Slice width 5.0 mm 1.0 mm
Feed/Rotation 9.0 mm
Rotation time 0.5 sec.
Kernel B31f B20f
Increment 5.0 mm 0.7 mm
CTDI
Vol
12.48 mGy
Effective dose Male: 3.9 mSv
Female: 4.7 mSv
Page view 295