Children
HeadRoutine05s
Indications:
Spiral mode for routine head studies, e.g. tumors,
hydrocephalus, hemorrhaging, abnormalities, etc.
A typical range of 12 cm covered in 5.1 sec.
Contrast medium IV injection
Start delay exam dependent
Flow rate dependent upon needle size/Access site
Total amount 1 – 2 ml per kg of body weight
* The conversion factor for an 8-week-old, and a scan
range of 90 mm was used.
HeadRoutine05s
kV 120
Effective mAs 150
Rotation time 0.5 sec.
Slice collimation 1.5 mm
Slice width 4.0 mm
Feed/Rotation 13.1 mm
Increment 4.0 mm
Kernel C30f
CTDI
Vol
28.5 mGy
Effective dose Male: 2.3 mSv*
Female: 2.3 mSv*